• Title of article

    Comparative measurements between a new logging tool and a reference instrument

  • Author/Authors

    Locatelli، نويسنده , , Marcel and Mathieu، نويسنده , , Hervè and Bobroff، نويسنده , , Serge and Guillot، نويسنده , , Geneviève and Zinszner، نويسنده , , Bernard، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    4
  • From page
    593
  • To page
    596
  • Abstract
    Measurements were performed on reference samples (D2O-H2O mixtures) and on highly heterogeneous rocks (Vosges sandstone) with a new logging tool designed to give access to a high spatial resolution, below 1.5 cm on the vertical scale, for a toroidal sensitive volume of 20 cm3. The results were compared to measurements obtained on a clinical magnetic resonance imaging (MRI) equipment working at the same frequency (4.3 MHz). T2 differences as high as 30% were observed for the reference samples; the shortest values were obtained with the logging tool. Porosity profiles of the rock samples were also compared to reference profiles obtained with a conventional computed tomography (CT) scanner. Both nuclear magnetic resonance (NMR) measurements underevaluate porosity by 2–4% for short T2 values (<10 ms).
  • Keywords
    Nuclear magnetic resonance logging tool , MAGNETIC RESONANCE IMAGING , Porous media
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    1998
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1828065