Title of article :
Comparative measurements between a new logging tool and a reference instrument
Author/Authors :
Locatelli، نويسنده , , Marcel and Mathieu، نويسنده , , Hervè and Bobroff، نويسنده , , Serge and Guillot، نويسنده , , Geneviève and Zinszner، نويسنده , , Bernard، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
4
From page :
593
To page :
596
Abstract :
Measurements were performed on reference samples (D2O-H2O mixtures) and on highly heterogeneous rocks (Vosges sandstone) with a new logging tool designed to give access to a high spatial resolution, below 1.5 cm on the vertical scale, for a toroidal sensitive volume of 20 cm3. The results were compared to measurements obtained on a clinical magnetic resonance imaging (MRI) equipment working at the same frequency (4.3 MHz). T2 differences as high as 30% were observed for the reference samples; the shortest values were obtained with the logging tool. Porosity profiles of the rock samples were also compared to reference profiles obtained with a conventional computed tomography (CT) scanner. Both nuclear magnetic resonance (NMR) measurements underevaluate porosity by 2–4% for short T2 values (<10 ms).
Keywords :
Nuclear magnetic resonance logging tool , MAGNETIC RESONANCE IMAGING , Porous media
Journal title :
Magnetic Resonance Imaging
Serial Year :
1998
Journal title :
Magnetic Resonance Imaging
Record number :
1828065
Link To Document :
بازگشت