Title of article :
Calculation of electric fields induced near metal implants by magnetic resonance imaging switched-gradient magnetic fields
Author/Authors :
Buechler، نويسنده , , D.N. and Durney، نويسنده , , C.H. and Christensen، نويسنده , , D.A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
10
From page :
1157
To page :
1166
Abstract :
Electric (E) fields induced near metal implants by MRI switched-gradient magnetic fields are calculated by a new equivalent-circuit numerical technique. Induced E-field results are found for a metallic spinal-fusion implant consisting of two thin wires connected to the metallic case of a current generator as well as for its subsections: a bare U-shaped wire, an insulated U-shaped wire, a cut insulated wire, and a generator. The presence of the metallic implants perturbs the E field significantly. Near the ends of the bare U-shaped wire, the E field is 89.7 times larger than in the absence of the wire. The greatest E field concentration occurs near the ends of the cut insulated wire, where the E field is 196.7 times greater than in the absence of the wire. In all cases, the perturbation of the induced E field by the implanted wire is highly localized within a few diameters of the wire.
Keywords :
Magnetic Resonance Imaging , Metallic implants , Induced electric fields , Safety
Journal title :
Magnetic Resonance Imaging
Serial Year :
1997
Journal title :
Magnetic Resonance Imaging
Record number :
1829554
Link To Document :
بازگشت