Title of article :
The effects of K-space data undersampling and discontinuities in keyhole functional MRI
Author/Authors :
Xiong، نويسنده , , Jinhu and Fox، نويسنده , , Peter T. and Gao، نويسنده , , Jia-Hong، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
The effects of the k-space data undersampling and the discontinuities associated with the keyhole approach in functional magnetic resonance imaging (fMRI) have been addressed. Undersampling of high k-space data increases the high spatial-frequency noise by a factor of m, where m is the number of images sharing the high k-space data. However, the effects of the noise can be effectively removed using clustered-pixel statistical analysis. The amplitude and phase discontinuities may result in an increase in the low spatial-frequency noise level, which has no impact on specificity and only a minimal impact on sensitivity. Although dramatically reducing acquisition time and preserving the spatial resolution of functional magnetic resonance imaging, the keyhole technique may have a limited capability in detecting highly focal activations.
Keywords :
Functional MRI , keyhole , functional imaging , Algorithms
Journal title :
Magnetic Resonance Imaging
Journal title :
Magnetic Resonance Imaging