Title of article :
High-pressure magnetic resonance imaging up to 40 MPa
Author/Authors :
Hirai، نويسنده , , Shuichiro and Kuwano، نويسنده , , Kunihiro and Ogawa، نويسنده , , Kuniyasu and Iriguchi، نويسنده , , Norio and Okazaki، نويسنده , , Ken، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
5
From page :
221
To page :
225
Abstract :
A high-pressure vessel designed for use with commercial magnetic resonance imaging equipment at up to 40 MPa of pressure was used and tested. Special features of the vessel are the following: 1) 12.6 mm sample chamber i.d.; 2) only non-magnetic parts; 3) visible sample from the outside; 4) resistant to corrosive chemicals, and; 5) sample could be manually translated and rotated in situ. This apparatus was demonstrated through observation of CO2 clathrate-hydrate growth in a water droplet injected into liquid CO2 at 20 MPa.
Keywords :
Operating system , Corrosive sample , CO2 clathrate hydrate , High-pressure MRI
Journal title :
Magnetic Resonance Imaging
Serial Year :
2000
Journal title :
Magnetic Resonance Imaging
Record number :
1830528
Link To Document :
بازگشت