Title of article :
High-resolution diffusion imaging using phase-corrected segmented echo-planar imaging
Author/Authors :
Brockstedt، نويسنده , , Sara and Moore، نويسنده , , James R and Thomsen، نويسنده , , Carsten and Holtهs، نويسنده , , Stig and Stهhlberg، نويسنده , , Freddy، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
9
From page :
649
To page :
657
Abstract :
Diffusion magnetic resonance imaging (MRI) was performed with a high-resolution segmented echo-planar imaging technique, which provided images with substantially less susceptibility artifacts than images obtained with single-shot echo-planar imaging (EPI). Diffusion imaging performed with any multishot pulse sequence is inherently sensitive to motion artifacts and in order to reduce motion artifacts, the presented method utilizes navigator echo phase corrections, performed after a one-dimensional Fourier transform along the frequency-encoding direction. Navigator echo phases were fitted to a straight line prior to phase correction to avoid errors from internal motion. In vivo imaging was performed using electro cardiographic (ECG) triggering. Apparent diffusion coefficient (ADC) maps were calculated on a pixel-by-pixel basis using up to seven diffusion sensitivities, ranging from b = 0 to 1129 × 106 s/m2.
Keywords :
Diffusion imaging , high resolution , Phase corrections , Segmented EPI , Ischemic stroke
Journal title :
Magnetic Resonance Imaging
Serial Year :
2000
Journal title :
Magnetic Resonance Imaging
Record number :
1830681
Link To Document :
بازگشت