• Title of article

    MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging

  • Author/Authors

    Hedlund، نويسنده , , Laurence W and Cofer، نويسنده , , Gary P and Owen، نويسنده , , Steven J and Allan Johnson، نويسنده , , G، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    7
  • From page
    753
  • To page
    759
  • Abstract
    We describe an MR-compatible ventilator that is computer controlled to generate a variety of breathing patterns, to minimize image degrading effects of breathing motion, and to support delivery of gas anesthesia and experimental inhalational gases. A key feature of this ventilator is the breathing valve that attaches directly to the endotracheal tube to reduce dead volume and allows independent control of inspiratory and expiratory phases of ventilation. This ventilator has been used in a wide variety of MR and x-ray microscopy studies of small animals, especially for MR imaging the lungs with hyperpolarized gases (3He & 129Xe).
  • Keywords
    MR microscopy , respirator , Mechanical Ventilation , Rodents , Xenon , Helium , Hyperpolarized gas
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2000
  • Journal title
    Magnetic Resonance Imaging
  • Record number