Title of article :
Special designed RF-antenna with integrated non-invasive carbon electrodes for simultaneous magnetic resonance imaging and electroencephalography acquisition at 7T
Author/Authors :
Van Audekerkea، نويسنده , , J and Peeters، نويسنده , , R and Verhoye، نويسنده , , M and Sijbers، نويسنده , , J and Van der Linden، نويسنده , , A، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
5
From page :
887
To page :
891
Abstract :
The construction of a high quality MR RF-antenna with incorporated EEG electrodes for simultaneous MRI and EEG acquisition is presented. The antenna comprises an active decoupled surface coil for receiving the MR signal and a whole body coil for transmitting the excitation RF pulses. The surface coil offers a high signal-to-noise ratio required for fMRI application and the whole body coil has a good B1 excitation profile, which enables the application of homogeneous RF pulses. Non-invasive carbon electrodes are used in order to minimise the magnetic susceptibility artefacts that occur upon application of conductive materials. This dedicated set-up is compared to a standard set-up being a linear birdcage coil and commercially available Ag/AgCl electrodes. As the acquired EEG signals are heavily disturbed by the gradient switching, intelligent filtering is applied to obtain a clean EEG signal.
Keywords :
RF-antenna , MRI , EEG
Journal title :
Magnetic Resonance Imaging
Serial Year :
2000
Journal title :
Magnetic Resonance Imaging
Record number :
1830779
Link To Document :
بازگشت