Title of article
The quantitative 19F-imaging of albumin at 1.5 T: a potential in-vivo tool
Author/Authors
Andrzej Dzik-Jurasz، نويسنده , , A.S.K and Wolber، نويسنده , , J and Prock، نويسنده , , T and Collins، نويسنده , , D.J and Leach، نويسنده , , M.O and Rowland، نويسنده , , I.J، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
6
From page
839
To page
844
Abstract
19F-MR-imaging has been used to quantitate albumin concentration in a phantom at 1.5 T. The experimentally derived relationship between albumin concentration and the T1 relaxation time of a fluorinated marker, tetrafluorosuccinic acid (TFSA) was used to calculate the albumin concentration from a quantitative 19F T1 map acquired using a gradient echo sequence. There was close correlation between calculated and actual BSA concentrations (r = 0.99, SE = 0.15). The potentially interfering effect of paramagnetic species on T1 relaxation times was also investigated. Relaxivity data show that albumin concentration measurements should be performed prior to any contrast agent administration.
Keywords
Nuclear magnetic resonance , Albumin , fluorine , tetrafluorosuccinic acid