    On the use of the FLAIR technique to improve the correction of eddy current induced artefacts in MR diffusion tensor imaging

  • Author/Authors

    Bastin، نويسنده , , Mark E.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    14
  • From page
    937
  • To page
    950
  • Abstract
    Eddy current induced geometric distortions can only be accurately corrected in brain diffusion-weighted echo-planar (DW-EP) images for b-values less than approximately 300 s mm−2 using the iterative cross-correlation (ICC) algorithm. This is due to the difference in signal intensity of the cerebrospinal fluid (CSF) compartment in the diffusion-weighted and baseline T2-weighted echo-planar (T2W-EP) images. At larger values of b, image misalignment artefacts can, however, be removed by directly correlating CSF-suppressed T2W-EP images with non-CSF-suppressed and CSF-suppressed DW-EP images. Separate phantom experiments can also be performed to provide eddy current calibration data. Here the ability of these methods to remove eddy current induced artefacts from DW-EP images collected in volunteer diffusion tensor imaging (DTI) experiments is investigated. Monte Carlo simulations show that in order for the ICC algorithm to produce accurate estimates of the eddy current induced distortions at b-values greater than 1000 s mm−2, the degree of CSF suppression should be greater than approximately 80%. This condition is typically met for FLAIR inversion times between 0.5 and 0.8 of the spin-lattice relaxation time of CSF. In volunteer studies the most complete image realignment was provided by direct correlation of CSF-suppressed T2W-EP and DW-EP images acquired in the FLAIR DTI experiment. These results indicate that although calibration data obtained from brain or phantom images can significantly reduce eddy current induced distortions, the optimum image realignment achievable using post-processing methods is likely to be that obtained by direct image warping techniques.
  • Keywords
    Warping , FLAIR , diffusion , Echo-planar , distortion
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2001
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1831126