• Title of article

    Elimination of k-space spikes in fMRI data

  • Author/Authors

    Zhang، نويسنده , , Xiaodong and Van De Moortele، نويسنده , , Pierre-Francois and Pfeuffer، نويسنده , , Josef and Hu، نويسنده , , Xiaoping، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    5
  • From page
    1037
  • To page
    1041
  • Abstract
    The subtle signal changes in functional magnetic resonance imaging (fMRI) can be easily overwhelmed by noise of various origins. Spikes in the collected fMRI raw data often arise from high-duty usage of the scanner hardware and can introduce significant noise in the image and thereby in the image time series. Consequently, the spikes will corrupt the functional data and degrade the result of functional mapping. In this work, a simple method based on processing the time course of the k-space data are introduced and implemented to remove the spikes in the acquired data. Application of the method to experimental data shows that the methods are robust and effective for eliminating of spike-related noise in fMRI time series.
  • Keywords
    EPI , k-Space , Spike noise removal , FMRI
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2001
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1831153