• Title of article

    Rapid and accurate measurement of transverse relaxation times using a single shot multi-echo echo-planar imaging sequence

  • Author/Authors

    Tyler، نويسنده , , Damian J and Moore، نويسنده , , Rachel J and Marciani، نويسنده , , Luca and Gowland، نويسنده , , Penny A، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    7
  • From page
    1031
  • To page
    1037
  • Abstract
    Methods for making rapid and accurate measurements and maps of the transverse relaxation time from a single free induction decay (FID) are proposed. The methods use a multi-echo sequence in combination with B1 insensitive (hyperbolic secant or BIREF2b) refocusing pulses and rapid echo-planar imaging techniques. The results were calibrated against a single spin echo echo-planar imaging sequence using a phantom containing a range of CuSO4 concentrations. The mean percentage absolute difference between the multi-echo and single-echo results was 3% for the multi-echo sequence using the hyperbolic secant refocusing pulse, and 7% for the multi-echo sequence using the BIREF2b refocusing pulse, compared to 13% for a multi-echo sequence using a nonselective sinc refocusing pulse. The use of the sequences in vivo has been demonstrated in studies of gastric function, i.e., the measurement of gastric dilution and monitoring of formation of a raft of alginate polysaccharide within the stomach.
  • Keywords
    Rapid measurement , quantitation , Transverse relaxation , EPI
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2004
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1832009