Title of article :
Estimation of the content of fat and parenchyma in breast tissue using MRI T1 histograms and phantoms
Author/Authors :
Boston، نويسنده , , Raymond C. and Schnall، نويسنده , , Mitchell D. and Englander، نويسنده , , Sarah A. and Landis، نويسنده , , J. Richard and Moate، نويسنده , , Peter J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
Mammographic breast density has been correlated with breast cancer risk. Estimation of the volumetric composition of breast tissue using three-dimensional MRI has been proposed, but accuracy depends upon the estimation methods employed. The use of segmentation based on T1 relaxation rates allows quantitative estimates of fat and parenchyma volume, but is limited by partial volume effects. An investigation employing phantom breast tissue composed of various combinations of chicken breast (to represent parenchyma) and cooking fats was carried out to elucidate the factors that influence MRI T1 histograms. Using the phantoms, T1 histograms and their known fat and parenchyma composition, a logistic distribution function was derived to describe the apportioning of the T1 histogram to fat and parenchyma. This function and T1 histograms were then used to predict the fat and parenchyma content of breasts from 14 women. Using this method, the composition of the breast tissue in the study population was as follows: fat 69.9±22.9% and parenchyma 30.1±22.9%.
Keywords :
MRI , breast tissue , FAT , Parenchyma , Estimation
Journal title :
Magnetic Resonance Imaging
Journal title :
Magnetic Resonance Imaging