Title of article
An investigation of pipeline materials for continuous hyperpolarized 129Xe gas spectroscopy
Author/Authors
Saito، نويسنده , , Moyoko and Hiraga، نويسنده , , Takashi and Hattori، نويسنده , , Mineyuki and Murakami، نويسنده , , Shumei and Nakai، نويسنده , , Toshiharu، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
4
From page
607
To page
610
Abstract
In order to establish a continuous hyperpolarized xenon-129 (HP-129Xe) gas delivery system for MR imaging, the effect of the metallic materials in the gas pipeline on the signal intensity was investigated. In the gas pipeline, an appropriate surface is needed to minimize wall relaxation by the HP-129Xe gas caused by the interaction between the HP gas and the surface, which can lead to signal loss. Although Pyrex glass is a popular material for the HP gas chamber, it is fragile under heat or physical stress.
s study, five stainless steel tubes (STs) prepared with different surface film-forming processes were examined. The MR signal intensities of HP-129Xe gas that passed through each tube were then compared. The film passivated by iron fluoride maintained the highest level of hyperpolarization, whereas that passivated by chromium oxide maintained the lowest. A ST with an appropriate passive film may be a useful alternative to a Pyrex glass pipeline.
Keywords
Hyperpolarized gas , magnetic resonance spectroscopy (MRS) , Continuous gas delivery system , 129Xe , Gas Pipeline