• Title of article

    Comparison of different methods for combining phase-contrast images obtained with multiple coils

  • Author/Authors

    Thunberg، نويسنده , , Per and Karlsson، نويسنده , , Matts and Wigstrِm، نويسنده , , Lars، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    795
  • To page
    799
  • Abstract
    The ability to determine coil sensitivities implies that a method optimized in terms of maximized signal-to-noise ratio (SNR) can be applied to the combination of multiple coil images. An optimization of SNR subsequently results in a minimized variance in quantitative velocity measurements using phase-contrast imaging. When coil sensitivities are unknown, the weighted mean method, utilizing the square of the signal magnitude as weights, is suitable for combination of multiple phase images. In this study, the optimized method using estimated coil sensitivities was compared to the weighted mean method both theoretically and experimentally. It is shown that absence of noise correlation between the different coil images implies no difference between the methods regarding the variance of the phase. In the practical situation, noise correlation does exist, implying an opportunity for further reduction of phase variance using the optimized method. In vitro and in vivo studies showed, however, no significant difference between the two methods studied.
  • Keywords
    Phase-contrast imaging , Reproducibility , sense , Parallel MRI
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2005
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1832175