Title of article :
Investigations of silicone breast implants with the NMR-MOUSE
Author/Authors :
Krüger، نويسنده , , Mirko and Schwarz، نويسنده , , Annett and Blümich، نويسنده , , Bernhard، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
4
From page :
215
To page :
218
Abstract :
Silicone breast implants are used for breast augmentation and breast reconstruction. The issues of concern associated with such implants are: (a) the quality control of each implant before implantation, and (b) the detection of implant bleeding after implantation. We have studied the use of the Nuclear Magnetic Resonance–MObile Universal Surface Explorer (NMR-MOUSE) for the nondestructive testing of (a) the quality of implant shells, and (b) changes in implant gel due to leakage of body fluid into the implant. Depth profiles measured nondestructively through implant shells at different positions of each implant by the Profile NMR-MOUSE assured good reproducibility of the quality and thickness of different shell layers. The leakage of implants upon rupture was mimicked by observing changes in the transverse NMR relaxation time of the implant gel upon ingress of physiological saline solution and safflower oil through the rupture. Results demonstrate that nondestructive testing with unilateral NMR is a potential method for use in the quality control of implants and for the screening of implants for rupture after implantation.
Keywords :
Mobile NMR , NMR profiling , NMR-MOUSE , Single-sided NMR
Journal title :
Magnetic Resonance Imaging
Serial Year :
2007
Journal title :
Magnetic Resonance Imaging
Record number :
1832401
Link To Document :
بازگشت