Title of article :
Simulation study on active noise control for a 4-T MRI scanner
Author/Authors :
Li، نويسنده , , Mingfeng and Lim، نويسنده , , Teik C. and Lee، نويسنده , , Jing-Huei، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
8
From page :
393
To page :
400
Abstract :
The purpose of this work is to study computationally the possibility of the application of a hybrid active noise control technique for magnetic resonance imaging (MRI) acoustic noise reduction. id control system combined with both feedforward and feedback loops embedded is proposed for potential application on active MRI noise reduction. A set of computational simulation studies were performed. Sets of MRI acoustic noise emissions measured at the patientʹs left ear location were recorded and used in the simulation study. paring three different control systems, namely, the feedback, the feedforward and the hybrid control, our results revealed that the hybrid control system is the most effective. The hybrid control system achieved approximately a 20-dB reduction at the principal frequency component. cluded that the proposed hybrid active control scheme could have a potential application for MRI scanner noise reduction.
Keywords :
active noise control , MRI , Acoustic noise
Journal title :
Magnetic Resonance Imaging
Serial Year :
2008
Journal title :
Magnetic Resonance Imaging
Record number :
1832675
Link To Document :
بازگشت