• Title of article

    Double-spin-echo diffusion weighting with a modified eddy current adjustment

  • Author/Authors

    Finsterbusch، نويسنده , , Jürgen، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    7
  • From page
    434
  • To page
    440
  • Abstract
    Magnetic field inhomogeneities like eddy current-related gradient fields cause geometric distortions in echo-planar imaging (EPI). This in particular affects diffusion-weighted imaging where these distortions vary with the direction of the diffusion weighting and hamper the accurate determination of diffusion parameters. The double-spin-echo preparation often used aims to reduce the cumulative eddy current effect by adjusting the diffusion-weighting gradient pulse durations to the time constant of the dominant eddy current contribution. However, eddy currents with a variety of time constants may be present and cause residual distortions. Here, a modification is proposed where the two bipolar gradient pairs of the preparation are adjusted independently to different time constants. At the expense of a slightly prolonged echo time, residual geometric distortions and correspondingly increased values of the diffusion anisotropy can be reduced as is demonstrated in phantoms and the human brain. Thus, it may help to improve the reliability of diffusion-weighted EPI.
  • Keywords
    Diffusion weighting , EPI , Geometric distortions , Eddy-current compensation , Eddy currents
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2010
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1832975