“MR conditional” respiratory ventilator system incident in a 3-T MRI environment
Author/Authors
Greenberg، نويسنده , , Kathryn L.Z. and Weinreb، نويسنده , , Jeffrey and Shellock، نويسنده , , Frank G.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
5
From page
1150
To page
1154
Abstract
The misunderstanding of the labeling for an “MR conditional” respiratory ventilator system resulted in the placement of this device, which had substantial ferromagnetic components, too close to a 3-T magnetic resonance (MR) system, causing a projectile incident. Magnetic resonance imaging health care professionals should be aware of the potentially dangerous consequences when such medical devices that have Food and Drug Administration-approved MR conditional labeling are brought into the MR system room. Recommendations to prevent future incidents are provided herein.
Keywords
Magnetic Resonance Imaging , MRI , MRI safety , accidents