• Title of article

    “MR conditional” respiratory ventilator system incident in a 3-T MRI environment

  • Author/Authors

    Greenberg، نويسنده , , Kathryn L.Z. and Weinreb، نويسنده , , Jeffrey and Shellock، نويسنده , , Frank G.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    5
  • From page
    1150
  • To page
    1154
  • Abstract
    The misunderstanding of the labeling for an “MR conditional” respiratory ventilator system resulted in the placement of this device, which had substantial ferromagnetic components, too close to a 3-T magnetic resonance (MR) system, causing a projectile incident. Magnetic resonance imaging health care professionals should be aware of the potentially dangerous consequences when such medical devices that have Food and Drug Administration-approved MR conditional labeling are brought into the MR system room. Recommendations to prevent future incidents are provided herein.
  • Keywords
    Magnetic Resonance Imaging , MRI , MRI safety , accidents
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2011
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1833214