Title of article :
A 3D balanced-SSFP Dixon technique with group-encoded k-space segmentation for breath-held non–contrast-enhanced MR angiography
Author/Authors :
Saranathan، نويسنده , , Manojkumar and Bayram، نويسنده , , Ersin and Worters، نويسنده , , Pauline W. and Glockner، نويسنده , , James F.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Abstract :
A three-dimensional balanced steady-state free precession (b-SSFP)-Dixon technique with a novel group-encoded k-space segmentation scheme called GUINNESS (Group-encoded Ungated Inversion Nulling for Non-contrast Enhancement in the Steady State) was developed. GUINNESS was evaluated for breath-held non–contrast-enhanced MR angiography of the renal arteries on 18 subjects (6 healthy volunteers, 12 patients) at 3.0 T. The method provided high signal-to-noise and contrast renal angiograms with homogeneous fat and background suppression in short breath-holds on the order of 20 s with high spatial resolution and coverage. GUINNESS has potential as a short breath-hold alternative to conventional respiratory-gated methods, which are often suboptimal in pediatric subjects and patients with significant diaphragmatic drift/sleep apnea.
Keywords :
Non–contrast-enhanced MR angiography , Balanced steady-state free precession , Dixon fat–water separation
Journal title :
Magnetic Resonance Imaging
Journal title :
Magnetic Resonance Imaging