• Title of article

    Detecting cortical lesions in multiple sclerosis at 7 T using white matter signal attenuation

  • Author/Authors

    Bluestein، نويسنده , , Katharine T. and Pitt، نويسنده , , David and Sammet، نويسنده , , Steffen and Zachariah، نويسنده , , Cherian Renil and Nagaraj، نويسنده , , Usha and Knopp، نويسنده , , Michael V. and Schmalbrock، نويسنده , , Petra، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    9
  • From page
    907
  • To page
    915
  • Abstract
    Cortical lesions have recently been a focus of multiple sclerosis (MS) MR research. In this study, we present a white matter signal attenuating sequence optimized for cortical lesion detection at 7 T. The feasibility of white matter attenuation (WHAT) for cortical lesion detection was determined by scanning eight patients (four relapsing/remitting MS, four secondary progressive MS) at 7 T. WHAT showed excellent gray matter–white matter contrast, and cortical lesions were hyperintense to the surrounding cortical gray matter, The sequence was then optimized for cortical lesion detection by determining the set of sequence parameters that produced the best gray matter–cortical lesion contrast in a 10-min scan. Despite the B1 inhomogeneities common at ultra-high field strengths, WHAT with an adiabatic inversion pulse showed good cortical lesion detection and would be a valuable component of clinical MS imaging protocols.
  • Keywords
    Cortical lesions , MRI , 7 T , optimization , white matter
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2012
  • Journal title
    Magnetic Resonance Imaging
  • Record number