Title of article :
Effects of RF profile on precision of quantitative T2 mapping using dual-echo steady-state acquisition
Author/Authors :
Wu، نويسنده , , Pei-Hsin and Cheng، نويسنده , , Cheng-Chieh and Wu، نويسنده , , Ming-Long and Chao، نويسنده , , Tzu-Cheng and Chung، نويسنده , , Hsiao-Wen and Huang، نويسنده , , Teng-Yi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Pages :
5
From page :
102
To page :
106
Abstract :
The dual echo steady-state (DESS) sequence has been shown successful in achieving fast T2 mapping with good precision. Under-estimation of T2, however, becomes increasingly prominent as the flip angle decreases. In 3D DESS imaging, therefore, the derived T2 values would become a function of the slice location in the presence of non-ideal slice profile of the excitation RF pulse. Furthermore, the pattern of slice-dependent variation in T2 estimates is dependent on the RF pulse waveform. Multi-slice 2D DESS imaging provides better inter-slice consistency, but the signal intensity is subject to integrated effects of within-slice distribution of the actual flip angle. Consequently, T2 measured using 2D DESS is prone to inaccuracy even at the designated flip angle of 90°. In this study, both phantom and human experiments demonstrate the above phenomena in good agreement with model prediction.
Keywords :
Dual-echo steady-state imaging , T2 quantification , Flip angle dependency , 2D and 3D imaging , Slice profile