A new vascular coupling device: Assessment of MRI issues at 3-tesla
Author/Authors :
Titterington، نويسنده , , Blake and Puschmann، نويسنده , , Carmen and Shellock، نويسنده , , Frank G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Pages :
5
From page :
585
To page :
589
Abstract :
AbstractObjective ar grafting frequently involves a time-consuming operation. A new vascular coupling device (VCD) made from metallic material was recently developed that may be advantageous because of the reduced operative time and decreased patient risks. Because of the metal, there are safety concerns related to MRI. Therefore, the purpose of this investigation was to use standardized testing techniques to evaluate MRI issues for this VCD in association with a 3-Tesla MR system. s D (corlife oHG, Hannover, Germany) was evaluated for magnetic field interactions (translational attraction and torque), MRI-related heating, and artifacts at 3-Tesla. MRI-related heating was assessed with the VCD in a gelled-saline-filled phantom with MRI performed at a whole body averaged SAR of 2.9-W/kg for 15-min. Artifacts were assessed using T1-weighted, spin echo, and gradient echo pulse sequences. s D exhibited minor magnetic field interactions and minimal heating (maximum temperature elevation, 1.8 °C). Artifacts were relatively small in relation to the size and shape of this implant. The lumen of the VCD could not be visualized using the gradient echo pulse sequence. sions tallic VCD that underwent evaluation is MR conditional for a patient undergoing an MRI procedure at 3-Tesla or less.
Keywords :
MRIsafety , MRIImplants , specific absorption rate , Artifacts
Journal title :
Magnetic Resonance Imaging
Serial Year :
2014
Journal title :
Magnetic Resonance Imaging
Record number :
1834219
Link To Document :
بازگشت