• Title of article

    3D high-resolution contrast enhanced MRI of carotid atheroma — a technical update

  • Author/Authors

    Zhu، نويسنده , , Chengcheng and Sadat، نويسنده , , Umar and Patterson، نويسنده , , Andrew J. and Teng، نويسنده , , Zhongzhao and Gillard، نويسنده , , Jonathan H. and Graves، نويسنده , , Martin J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2014
  • Pages
    4
  • From page
    594
  • To page
    597
  • Abstract
    AbstractObjective pment of a fast 3D high-resolution magnetic resonance imaging (MRI) protocol for improved carotid artery plaque imaging. s tients with carotid atherosclerosis disease underwent 3D high-resolution MRI which included time-of-flight and T1-weighted variable flip angle, fast-spin-echo (FSE) imaging, pre- and post-intravenous gadolinium-based contrast agent administration. s uality images with intrinsic blood suppression were obtained pre- and post-contrast administration using a 3D FSE sequence. The plaque burden, lipid core volume, hemorrhage volume and fibrous cap thickness were well determined. sions h-resolution MR imaging of carotid plaque using TOF and 3D FSE can achieve high isotropic resolution, large coverage, and excellent image quality within a short acquisition time.
  • Keywords
    Carotid atheroma , 3D Fast Spin Echo , Hemorrhage , Lipid core , Fibrous cap
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2014
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1834227