• Title of article

    Maximizing MR signal for 2D UTE slice selection in the presence of rapid transverse relaxation

  • Author/Authors

    Carl، نويسنده , , Michael and Chiang، نويسنده , , Jing-Tzyh Alan and Du، نويسنده , , Jiang، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2014
  • Pages
    6
  • From page
    1006
  • To page
    1011
  • Abstract
    Ultrashort TE (UTE) sequences allow direct visualization of tissues with very short T2 relaxation times, such as tendons, ligaments, menisci, and cortical bone. In this work, theoretical calculations, simulations, and phantom studies, as well as in vivo imaging were performed to maximize signal-to-noise ratio (SNR) for slice selective RF excitation for 2D UTE sequences. The theoretical calculations and simulations were based on the Bloch equations, which lead to analytic expressions for the optimal RF pulse duration and amplitude to maximize magnetic resonance signal in the presence of rapid transverse relaxation. In steady state, it was found that the maximum signal amplitude was not obtained at the classical Ernst angle, but at an either lower or higher flip angle, depending on whether the RF pulse duration or amplitude was varied, respectively.
  • Keywords
    RF Optimization , 2D UTE , Ernst angle , Short T2
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2014
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1834472