Title of article
Maximizing MR signal for 2D UTE slice selection in the presence of rapid transverse relaxation
Author/Authors
Carl، نويسنده , , Michael and Chiang، نويسنده , , Jing-Tzyh Alan and Du، نويسنده , , Jiang، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2014
Pages
6
From page
1006
To page
1011
Abstract
Ultrashort TE (UTE) sequences allow direct visualization of tissues with very short T2 relaxation times, such as tendons, ligaments, menisci, and cortical bone. In this work, theoretical calculations, simulations, and phantom studies, as well as in vivo imaging were performed to maximize signal-to-noise ratio (SNR) for slice selective RF excitation for 2D UTE sequences. The theoretical calculations and simulations were based on the Bloch equations, which lead to analytic expressions for the optimal RF pulse duration and amplitude to maximize magnetic resonance signal in the presence of rapid transverse relaxation. In steady state, it was found that the maximum signal amplitude was not obtained at the classical Ernst angle, but at an either lower or higher flip angle, depending on whether the RF pulse duration or amplitude was varied, respectively.
Keywords
RF Optimization , 2D UTE , Ernst angle , Short T2
Journal title
Magnetic Resonance Imaging
Serial Year
2014
Journal title
Magnetic Resonance Imaging
Record number
1834472
Link To Document