• Title of article

    Lʹoption scanneur sur simulateur: 14 mois dʹutilisation dans le service de radiothérapie du CHR dʹOrléans

  • Author/Authors

    Bouscayrol، نويسنده , , H and Lescrainier، نويسنده , , H. and Sabattier، نويسنده , , R and Wachter، نويسنده , , T and Breteau، نويسنده , , N، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    6
  • From page
    567
  • To page
    572
  • Abstract
    Summary rpose was to evaluate the use of the CT option for simulator during a 14-month period in the radiotherapy department. The CT option has been adapted on the Philips simulator SL23. The virtual tunnel diameter is 92 cm which allows slice acquisition regardless of the contention device used. This system is connected to the treatment planning system through an Ethernet link. Three fields of view are available which cover the standard radiotherapy use. Four hundred and twentyseven patients benefited from this system over a 14-month period of use. The number of slices acquired per patient regularly increased. The use of the system was rapidly extended to all the standard treatments. Those slices were also used as additional information for these high technology treatments. The good quality of the images and the reliabiblity of this system involved a rapid integration in the treatment preparation procedure. It will not replace the use of the scanner but will significantly improve the treatment quality.
  • Keywords
    SIMULATION , plan de traitement en radiothérapie assisté par ordinateur , Tomographie , Computer simulation , computer-assisted , tomography , X-Ray Computed , radiotherapy planning
  • Journal title
    Cancer Radiotherapie
  • Serial Year
    1997
  • Journal title
    Cancer Radiotherapie
  • Record number