• Title of article

    Adaptation de lʹanalyse receiver operating characteristics pour lʹévaluation quantitative des plans de traitement en radiothérapie stéréotaxique

  • Author/Authors

    Platoni، Kalliopi نويسنده , , K. and Lefkopoulos، نويسنده , , D. and Grandjean، نويسنده , , P. and Schlienger، نويسنده , , M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    9
  • From page
    494
  • To page
    502
  • Abstract
    Summary finition of criteria and of a methodology dedicated to the quantitative evaluation of conformal stereotactic treatment plans is presented. We implemented the ‘Receiver Operating Characteristics’ (ROC) analysis, already used in medical imaging, for the quantitative evaluation of irradiation treatment plans. This implementation is based on data provided by dose-volume histograms (DVH). Three techniques, each one using a different dosimetric criterion, were defined for the choice of a reference isodose for a given treatment plan. We used this ROC analysis for the selection of the most conformai treatment plan and its reference isodose among the treatment plans proposed for one patient. This study revealed the interest of ROC analysis based on dose-volume histograms for the quantitative evaluation of treatment plans.
  • Keywords
    analyse ROC , évaluation quantitative , histogrammes dose-volume , dose-volume histograms , Radiothérapie en conditions stéréotaxiques , Quantitative Evaluation , Stereotactic radiation therapy , ROC analysis
  • Journal title
    Cancer Radiotherapie
  • Serial Year
    1999
  • Journal title
    Cancer Radiotherapie
  • Record number