Title of article :
A comparison of double-plate fixation methods for complex distal humerus fractures
Author/Authors :
Self، نويسنده , , James and Viegas، نويسنده , , Steven F. and Buford Jr.، نويسنده , , William L. and Patterson، نويسنده , , Rita M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
A modified method of fracture fixation of complex distal humeral fractures with medial and lateral plates and bolts was biomechanically tested and compared with previously described fixation techniques. Compressive stiffness coefficients were determined for three classes of fixation before and after fatigue cycling. This procedure was followed with compressive loading to failure. The results show that in the most unstable fracture type tested this new fixation method provides increased strength and stability. Early clinical follow-up examinations of patients treated with this technique show that this method is a reasonable fixation alternative for the complex distal humerus intercondylar fracture.
Journal title :
Journal of Shoulder and Elbow Surgery
Journal title :