Title of article :
Dissociation of modular humeral head components: A biomechanical and implant retrieval study
Author/Authors :
Blevins، نويسنده , , Field T. and Deng، نويسنده , , Xianghua and Torzilli، نويسنده , , Peter A. and Dines، نويسنده , , David and Warren، نويسنده , , Russell F.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Abstract :
In vivo dissociation of the Morse-taper of shoulder arthroplasty modular humeral components has been reported. The incidence of this complication appears to be approximately 1:1000. The objective of this study was to identify conditions that might affect the Morse-taper interface strength in humeral components. Mechanical tests were performed to load and dissociate humeral heads from the humeral stems (titanium). The effect of loading rate, load amplitude, and number of impactions was investigated. Dissociation force was measured after the taper was contaminated with water, oil, blood, and bone cement particles. The mean dissociation force after two impactions with a mallet was 2926 ± 955 N. Dissociation force was linearly proportional to impaction force. Repetitive loading beyond two impactions did not significantly increase taper strength. Contamination of the taper with as little as 0.4 ml of fluid could prevent fixation of the taper. (J Shoulder Elbow Surg 1997;6:113-24.)
Journal title :
Journal of Shoulder and Elbow Surgery
Journal title :
Journal of Shoulder and Elbow Surgery