Title of article :
A standard method of shoulder strength measurement for the Constant score with a spring balance
Author/Authors :
Bankes، نويسنده , , Marcus J.K. and Crossman، نويسنده , , John E. and Emery، نويسنده , , Roger J.H.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
6
From page :
116
To page :
121
Abstract :
The strength component of the Constant score has been criticized for its lack of a precisely defined measurement method. A series of experiments was performed to compare three different methods in normal and pathologic shoulders with the use of a standard test position. These were (1) the Isobex isometric dynamometer, (2) Constantʹs unsecured spring balance, and (3) a new modification in which the spring balance is fixed at one end and the reading is taken after 5 seconds of maximum effort. The results suggest that this simple modification with a low-cost spring balance can give similar values to those from the Isobex. The need for precision of terms and a definition of the method is discussed, and recommendations for the standardization of the many variables in making this measurement are made. (J Shoulder Elbow Surg 1998;7:116-21.)
Journal title :
Journal of Shoulder and Elbow Surgery
Serial Year :
1998
Journal title :
Journal of Shoulder and Elbow Surgery
Record number :
1863873