• Title of article

    Axillary nerve injury after thermal capsular shrinkage of the shoulder

  • Author/Authors

    Greis، نويسنده , , Patrick E. and Burks، نويسنده , , Robert T. and Schickendantz، نويسنده , , Mark S. and Sandmeier، نويسنده , , Robert، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    5
  • From page
    231
  • To page
    235
  • Abstract
    The purpose of this study was to evaluate 4 cases in which axillary nerve injury was observed after radiofrequency capsular shrinkage of the shoulder. These cases were used to evaluate the clinical circumstances under which axillary nerve injury occurred, the time frame over which these injuries recovered, and the ultimate outcome observed in these patients. The case histories of each of these 4 patients were carefully evaluated. In 2 of the 4 cases, purely sensory findings were found, and in 2 cases both sensory and motor injuries occurred. In each case the inferior axillary recess was treated using the Oratech Interventions Tac-S probe. The authors conclude that injury to the axillary nerve during this procedure is possible and postulate that heat penetration through the capsule to the nerve is the most likely cause of injury. (J Shoulder Elbow Surg 2001;10:231-5.)
  • Journal title
    Journal of Shoulder and Elbow Surgery
  • Serial Year
    2001
  • Journal title
    Journal of Shoulder and Elbow Surgery
  • Record number

    1865295