Title of article :
The effect of inaccurate volume calibrations on hydrogen uptake measured by the Sieverts method
Author/Authors :
Webb، نويسنده , , C.J. and Gray، نويسنده , , E.MacA.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Pages :
7
From page :
2168
To page :
2174
Abstract :
The Sieverts technique is a popular method of measuring the uptake of gas on a sample and is an important tool for the characterisation of hydrogen uptake by potential hydrogen storage materials. An analysis of the consequences of errors in the calibration of volumes in this technique has been made using hydrogen absorption on an AB2 alloy as an example. Trends in the error in the uptake as a function of errors in the volumes have been explained in terms of the equations for uptake. Significant advantages are found for calibration techniques for which the ratio of volumes is an intrinsic measurement.
Keywords :
Volume calibrations , Hydrogen storage , Sieverts
Journal title :
International Journal of Hydrogen Energy
Serial Year :
2014
Journal title :
International Journal of Hydrogen Energy
Record number :
1867141
Link To Document :
بازگشت