    Variation in the position of the elbow flexion axis after total joint replacement with three different prostheses

  • Author/Authors

    Ericson، نويسنده , , Anne and Stark، نويسنده , , André and Arndt، نويسنده , , Anton، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    8
  • From page
    760
  • To page
    767
  • Abstract
    The instantaneous flexion axes in rheumatoid patients treated with a Kudo (Biomet UK Ltd, Swindon, UK), Capitellocondylar (Johnson & Johnson Orthopaedics Inc, Raynham, MA), or GSB III (Zimmer GmbH, Winterthur, Switzerland) elbow prosthesis were calculated using radiostereometric analysis and compared with those previously found in the normal elbow in vivo. Variability in the flexion axis position was greater in the prosthetic elbows than reported for healthy subjects. This was especially evident for the unlinked Kudo prosthesis, with dispersions of the axes up to 84.3° in the frontal and 45.0° in the horizontal plane. Corresponding values for the unlinked Capitellocondylar prosthesis were 19.7° and 20.9°. The linked GSB III prosthesis was the most constrained, with dispersions within 15.4° in the frontal and 4.6° in the horizontal plane. The unlinked elbow prostheses displayed considerable axis dispersion under physiologic conditions. Implant design seems to be able to compensate for this to some extent, as indicated, for example, by the results of the Capitellocondylar prosthesis.