Title of article :
Optimal screw placement for base plate fixation in reverse total shoulder arthroplasty
Author/Authors :
DiStefano، نويسنده , , James Guido and Park، نويسنده , , Andrew Y. and Nguyen، نويسنده , , Thuc-Quyen D. and Diederichs، نويسنده , , Gerd and Buckley، نويسنده , , Jenni M. and Montgomery III، نويسنده , , William H.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
Hypothesis
ar cortical thickness has not been fully characterized from the perspective of determining optimal screw placement for securing the glenoid base plate in reverse shoulder arthroplasty.
als and methods
fresh frozen cadaveric scapulae underwent high resolution CT scans with 3-dimensional reconstructions and wall thickness analysis. Digital base plates were positioned and virtual screws were placed according to 2 scenarios: A - intraosseous through the entire course and exits a “safe region” with no known neurovascular structures; B - may leave and re-enter the bone and penetrates the thickest cortical region accessible regardless of adjacent structures.
s
enario A, the optimal screw configurations were: (superior screw) length = 35 mm, 9° superior, 2° posterior; (inferior screw-A) length = 34 mm, 16° inferior, 5° anterior; (inferior screw-B) length = 31 mm, 31 inferior, 4 posterior; (posterior screw) length 19 mm, 29° inferior, 3° anterior. For scenario B: (superior screw) length = 36 mm, 28° superior, 10° anterior; (inferior screw) length = 35 mm, 19° inferior, 4° anterior; (posterior screw) length 37 mm, 23° superior, 3° anterior. The anterior screw was consistent between scenarios A and B, averaged 29 mm in length and was directed 16° inferior and 14° posterior.
sion
r cortical regions were present in the lateral aspect of the suprascapular notch, scapular spine base, anterior/superior aspect of inferior pillar and junction of glenoid neck and scapular spine. Regions with high cortical thickness were accessible for both scenarios except for the posterior screw in scenario A.
Keywords :
screw placement , Base plate , 3-dimensional , reverse total shoulder arthroplasty , Scapula , Cortical thickness
Journal title :
Journal of Shoulder and Elbow Surgery
Journal title :
Journal of Shoulder and Elbow Surgery