Title of article :
Reliability and agreement between 2 strength devices used in the newly modified and standardized Constant score
Author/Authors :
Kristensen، نويسنده , , Morten Tange and Aagesen، نويسنده , , Maria and Hjerrild، نويسنده , , Signe and Lund Skov Larsen، نويسنده , , Pernille and Hovmand، نويسنده , , Bente and Ban، نويسنده , , Ilija، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
Hypothesis
w and standardized test protocol for the Constant score (CS) provides new methodology, but different devices are still used for shoulder strength testing. It was hypothesized that strength measurements using the IsoForceControl (IFC) dynamometer (MDS Medical Device Solutions, Oberburg, Switzerland) would provide results comparable with the IDO isometer (Innovative Design Orthopaedics, Redditch, UK).
als and methods
healthy subjects, aged 19 to 83 years, were studied, with 5 men and 5 women in each of 6 ten-year age groups. The IFC and IDO were used in randomized order with an 8-minute interval between testing. Subjects performed 3 successive trials with strong verbal encouragement, with 1 minute between trials. The best strength performance was used in the analysis. The rater and subjects were blinded to all results.
s
C produced 0.28-kg (0.62-lb) higher strength values on average than the IDO (P = .002). The intraclass correlation coefficient (ICC2,1) was 0.97 (95% confidence interval, 0.95-0.98), whereas the standard error of measurement and smallest real difference were 0.43 kg (0.95 lb) and 1.2 kg (2.63 lb), respectively. The total CS and strength reached mean values of 92.4 points (SD, 6.2 points) and 8.2 kg (SD, 2.6 kg) (18.0 lb [SD, 5.8 lb]), respectively, and were negatively associated with age (r > −0.407, P ≤ .001). The strength values decreased (P ≤ .001) by 1.3 CS points per decade, and women had strength values that were 8 CS points lower on average than those of men of the same age.
sions
lative (intraclass correlation coefficient) and absolute (standard error of measurement) reliability between the IFC and IDO is excellent, indicating that performances reported from settings using the IDO are comparable with those recorded with the IFC in other settings.
Keywords :
Shoulder , Constant score , Intrarater reliability , Standardization , Agreement , strength devices
Journal title :
Journal of Shoulder and Elbow Surgery