Author/Authors :
De Lurgio، نويسنده , , David B. and Sathavorn، نويسنده , , Charn and Mera، نويسنده , , Fernando De Leon Gonzalez ، نويسنده , , Angel and Walter، نويسنده , , Paul F. and Langberg، نويسنده , , Jonathan J.، نويسنده ,
Abstract :
In summary, ICD lead abrasions are frequent findings during ICD generator replacement, and may result in lead failure. Careful inspection, as well as repair or replacement, of severely abraided leads should be considered.