• Title of article

    Comparison of the sideris and Amplatzer septal occlusion devices

  • Author/Authors

    Walsh، نويسنده , , Kevin P and Tofeig، نويسنده , , Magdi and Kitchiner، نويسنده , , Denise J and Peart، نويسنده , , Ian and Arnold، نويسنده , , Robert، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    4
  • From page
    933
  • To page
    936
  • Abstract
    The results of transcatheter atrial septal defect (ASD) occlusion with 2 different devices (Sideris adjustable buttoned device vs Amplatzer Septal Occluder) were compared in 2 consecutive series of patients. Comparative outcomes were assessed by whether a device was implanted or not, by complications and fluoroscopy time of implantation, and by the incidence of residual shunting on transthoracic echocardiography at follow-up. The patient and defect characteristics were similar in both groups. Twenty-eight of 33 Sideris devices and 37 of 39 Amplatzer devices were implanted. The fluoroscopy time for the Amplatzer implants was 13.4 minutes (range 8 to 41) compared with 23.7 minutes (range 11 to 60.6) for the Sideris implants (p <0.001). The complete occlusion rate for the Amplatzer device was 93% compared with 44% for the Sideris device at 1 year (p <0.001). In conclusion, the Amplatzer device produces higher occlusion rates of ASDs with shorter fluoroscopy times.
  • Journal title
    American Journal of Cardiology
  • Serial Year
    1999
  • Journal title
    American Journal of Cardiology
  • Record number

    1890145