Author/Authors :
Kim، نويسنده , , Sang-Wook and Mintz، نويسنده , , Gary S. and Hong، نويسنده , , Young-Joon and Pakala، نويسنده , , Rajbabu and Park، نويسنده , , Kyung-Sook and Pichard، نويسنده , , Augusto D. and Satler، نويسنده , , Lowell F. and Kent، نويسنده , , Kenneth M. and Suddath، نويسنده , , William O. and Waksman، نويسنده , , Ron and Weissman، نويسنده , , Neil J.، نويسنده ,
Abstract :
Intravascular ultrasound (IVUS) is used before and after intervention and at follow-up to assess the quality of the acute result as well as the long-term effects of stent implantation. Virtual histology (VH) IVUS classifies tissue into fibrous and fibrofatty plaque, dense calcium, and necrotic core. Although most interventional procedures include stent implantation, VH IVUS classification of stent metal has not been validated. In this study, the VH IVUS appearance of acutely implanted stents was assessed in 27 patients (30 lesions). Most stent struts (80%) appeared white (misclassified as “calcium”) surrounded by red (misclassified as “necrotic core”); 2% appeared just white, and 17% were not detectable (compared with grayscale IVUS because of the software-imposed gray medial stripe). The rate of “white surrounded by red” was similar over the lengths of the stents; however, undetectable struts were mostly at the distal edges (31%). Quantitatively, including the struts within the regions of interest increased the amount of “calcium” from 0.23 ± 0.35 to 1.07 ± 0.66 mm2 (p <0.0001) and the amount of “necrotic core” from 0.59 ± 0.65 to 1.31 ± 0.87 mm2 (p <0.0001). Most important, because this appearance occurs acutely, it is an artifact, and the red appearance should not be interpreted as peristrut inflammation or necrotic core when it is seen at follow-up. In conclusion, acutely implanted stents have an appearance that can be misclassified by VH IVUS as “calcium with or without necrotic core.” It is important not to overinterpret VH IVUS studies of chronically implanted stents when this appearance is observed at follow-up. A separate classification for stent struts is necessary to avoid these misconceptions and misclassifications.