Title of article
Usefulness of a hand-carried cardiac ultrasound device to detect clinically significant valvular regurgitation in hospitalized patients
Author/Authors
Kobal، نويسنده , , Sergio L. and Tolstrup، نويسنده , , Kirsten and Luo، نويسنده , , Huai and Neuman، نويسنده , , Yoram and Miyamoto، نويسنده , , Takashi and Mirocha، نويسنده , , James and Naqvi، نويسنده , , Tasneem Z. and Siegel، نويسنده , , Robert J.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
4
From page
1069
To page
1072
Abstract
The accuracy of a hand-carried cardiac ultrasound (HCCU) device for the diagnosis of valvular regurgitations using color-flow Doppler was assessed. One hundred twenty hospitalized patients with at least mild valvular regurgitation by standard echocardiography were evaluated by 2 cardiologists using HCCU. The HCCU operator identified 99% of clinically significant valvular regurgitations and assessed the severity correctly in 83% of cases. For mild regurgitation, the HCCU sensitivity and specificity were 82% and 93%, respectively, resulting in correct assessments of severity in 71% of cases. HCCU used by trained cardiologists has high sensitivity for the detection and accurate assessment of the severity of clinically relevant valvular regurgitation.
Journal title
American Journal of Cardiology
Serial Year
2004
Journal title
American Journal of Cardiology
Record number
1897381
Link To Document