• Title of article

    Usefulness of echocardiographic tissue synchronization imaging to predict acute response to cardiac resynchronization therapy

  • Author/Authors

    Gorcsan III، نويسنده , , John and Kanzaki، نويسنده , , Hideaki and Bazaz، نويسنده , , Raveen and Dohi، نويسنده , , Kaoru and Schwartzman، نويسنده , , David، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    4
  • From page
    1178
  • To page
    1181
  • Abstract
    Echocardiographic tissue synchronization imaging (TSI) consists of color-coding time-to-peak tissue Doppler velocities. This study of 29 patients who underwent cardiac resynchronization therapy (CRT) demonstrated that differences in baseline time-to-speak velocities of opposing ventricular walls by TSI were greater in 15 patients, with an acute hemodynamic improvement. A ≥65 ms delay from the anterior septum to the posterior wall using the apical long-axis view had 87% sensitivity and 100% specificity for predicting an acute response. Although a subgroup without acute improvement had later decreases in end-systolic volume, suggesting that acute response underestimates long-term effects, TSI has potential to assist in guiding CRT.
  • Journal title
    American Journal of Cardiology
  • Serial Year
    2004
  • Journal title
    American Journal of Cardiology
  • Record number

    1897436