Title of article :
Comparison of Left Ventricular Outflow Geometry and Aortic Valve Area in Patients With Aortic Stenosis by 2-Dimensional Versus 3-Dimensional Echocardiography
Author/Authors :
Saitoh، نويسنده , , Takeji and Shiota، نويسنده , , Maiko and Izumo، نويسنده , , Masaki and Gurudevan، نويسنده , , Swaminatha V. and Tolstrup، نويسنده , , Kirsten and Siegel، نويسنده , , Robert J. and Shiota، نويسنده , , Takahiro، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
6
From page :
1626
To page :
1631
Abstract :
The present study sought to elucidate the geometry of the left ventricular outflow tract (LVOT) in patients with aortic stenosis and its effect on the accuracy of the continuity equation-based aortic valve area (AVA) estimation. Real-time 3-dimensional transesophageal echocardiography (RT3D-TEE) provides high-resolution images of LVOT in patients with aortic stenosis. Thus, AVA is derived reliably with the continuity equation. Forty patients with aortic stenosis who underwent 2-dimensional transthoracic echocardiography (2D-TTE), 2-dimensional transesophageal echocardiography (2D-TEE), and RT3D-TEE were studied. In 2D-TTE and 2D-TEE, the LVOT areas were calculated as π × (LVOT dimension/2)2. In RT3D-TEE, the LVOT areas and ellipticity ([diameter of the anteroposterior axis]/[diameter of the medial-lateral axis]) were evaluated by planimetry. The AVA is then determined using planimetry and the continuity equation method. LVOT shape was found to be elliptical (ellipticity of 0.80 ± 0.08). Accordingly, the LVOT areas measured by 2D-TTE (median 3.7 cm2, interquartile range 3.1 to 4.1) and 2D-TEE (median 3.7 cm2, interquartile range 3.1 to 4.0) were smaller than those by 3D-TEE (median 4.6 cm2, interquartile range 3.9 to 5.3; p <0.05 vs both 2D-TTE and 2D-TEE). RT3D-TEE yielded a larger continuity equation-based AVA (median 1.0 cm2, interquartile range 0.79 to 1.3, p <0.05 vs both 2D-TTE and 2D-TEE) than 2D-TTE (median 0.77 cm2, interquartile range 0.64 to 0.94) and 2D-TEE (median 0.76 cm2, interquartile range 0.62 to 0.95). Additionally, the continuity equation-based AVA by RT3D-TEE was consistent with the planimetry method. In conclusion, RT3D-TEE might allow more accurate evaluation of the elliptical LVOT geometry and continuity equation-based AVA in patients with aortic stenosis than 2D-TTE and 2D-TEE.
Journal title :
American Journal of Cardiology
Serial Year :
2012
Journal title :
American Journal of Cardiology
Record number :
1902326
Link To Document :
بازگشت