• Title of article

    Impact of Smart Infusion Technology on Administration of Anticoagulants (Unfractionated Heparin, Argatroban, Lepirudin, and Bivalirudin)

  • Author/Authors

    Fanikos، نويسنده , , John and Fiumara، نويسنده , , Karen and Baroletti، نويسنده , , Steve and Luppi، نويسنده , , Carol and Saniuk، نويسنده , , Catherine and Mehta، نويسنده , , Amar and Silverman، نويسنده , , Jon and Goldhaber، نويسنده , , Samuel Z.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    4
  • From page
    1002
  • To page
    1005
  • Abstract
    This study reviewed 863 alerts generated from the infusion of anticoagulants in 355 patients from October 2003 to January 2005. Alerts were generated by smart infusion technology pumps and recorded in the devices’ memory. The most common alerts were underdose alerts (59.8%), followed by overdose alerts (31.3%) and duplicate drug therapy alerts (8.9%). In response to the alerts, users’ most frequent action was to cancel (46.5%) or reprogram (43.1%) the infusions. The highest percentage of alerts occurred from 2 to 4 p.m. During the study, there were 4 infusion rate errors, compared with 15 in the immediately preceding 16-month period. In conclusion, smart infusion technology intercepted keypad entry errors, thereby reducing the likelihood of intravenous anticoagulant overdose or underdose. Dose or infusion rate programming during intravenous anticoagulation is an important targets for medication safety interventions.
  • Journal title
    American Journal of Cardiology
  • Serial Year
    2007
  • Journal title
    American Journal of Cardiology
  • Record number

    1902909