Hou، نويسنده , , Jingbo and Jia، نويسنده , , Haibo and Huang، نويسنده , , Xingtao and Yu، نويسنده , , Huai and Ren، نويسنده , , Xuefeng and Fang، نويسنده , , Yan and Han، نويسنده , , Zhigang and Yang، نويسنده , , Shuang and Meng، نويسنده , , Lingbo and Zhang، نويسنده , , Shaosong and Yu، نويسنده , , Bo Yoon Jang، نويسنده , , Ik-Kyung، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
6
From page :
1117
To page :
1122
Abstract :
The aim of this study was to evaluate neointimal coverage obtained using a new method of polytetrafluoroethylene-covered stent (PCS) implantation combined with underlying longer sirolimus-eluting stent (SES) implantation using optical coherence tomography. Nine patients were enrolled in this study, including patients with coronary artery perforations, original coronary aneurysms, and acquired coronary aneurysms after drug-eluting stent implantation. All patients were first treated with long SES implantation and then with focal PCS implantation. Postprocedural and follow-up angiographic and optical coherence tomographic examinations were performed in all patients, and intravascular ultrasound was performed in 5 patients. All patients were asymptomatic during follow-up, without recurrent angina. There was no stent-edge or stent-segment binary restenosis. Values of late loss for proximal SES segments, PCS segments, and distal SES segments were similar (0.09, 0.07, and 0.04 mm, respectively, p = 0.8113). The mean neointimal thickness of PCS was less than that of proximal and distal SES. However, no malapposed cross sections or uncovered cross sections were found in PCS segments compared with SES segments (p = 0.0011). In conclusion, the combination of PCS and underlying longer SES implantation can offer better angiographic follow-up results. High-resolution optical coherence tomography provided convincing proof of full neointimal coverage of PCS. This new method of combined PCS and SES implantation may be a better choice compared with direct PCS implantation in certain clinical settings.
Journal title :
American Journal of Cardiology
Serial Year :
2013
Journal title :
American Journal of Cardiology
Record number :
1903415
Link To Document :
بازگشت