• Title of article

    Development and validation of ethylenethiourea determination in foods using methanol-based extraction, solid-phase extraction cleanup and LC–MS/MS

  • Author/Authors

    Tran، نويسنده , , Kevin and Mactal، نويسنده , , Lucie P. and Cromer، نويسنده , , Michele R. and Vocque، نويسنده , , Ralph H. and Smith، نويسنده , , Robert E.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    3
  • From page
    340
  • To page
    342
  • Abstract
    To response to the need for a rapid, cost-effective, eco-friendly and efficient extraction process, a sensitive method for the determination of ethylenethiourea (ETU) in food matrices by high-performance liquid chromatography–mass spectrometry (HPLC–LC/MS) was developed and validated. ETU was extracted from food commodities with methanol, cleaned up by alumina-SPE column, and then determined by HPLC–MS/MS. The MS detection was operated in positive ionization mode. For confirmation of target compound, two precursor ion > product ion transitions were selected by multi-reaction monitoring mode (MRM). The method showed good linearity with correlation coefficient higher than 0.9950. Recoveries at three spiked levels (10, 50, 100 ng/g) in random selected food matrices were in range of 71–121% with RSDs not larger than 25%. The limit of quantitation for the analyte was estimated at 5 ng/g.
  • Keywords
    Ethylene-bis-dithiocarbamate , Imidazolidine-2-thione , LC–MS/MS , Ethylenethiourea , ETU , EBDCs
  • Journal title
    Food Chemistry
  • Serial Year
    2013
  • Journal title
    Food Chemistry
  • Record number

    1944415