Title of article :
Simultaneous determination of maltose and glucose using a dual-electrode flow injection system
Author/Authors :
Qu، نويسنده , , Hongbo and Zhang، نويسنده , , Xian-En and Zhang، نويسنده , , Shu-Zheng، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
6
From page :
187
To page :
192
Abstract :
A dual-electrode system consisting of a glucose oxidase (GOD) electrode and an amyloglucosidase/glucose oxidase (AG) electrode in an FIA manifold was constructed to determine maltose and glucose simultaneously in their mixture. Highly active enzyme membranes were acquired. The linear range of the system was 0·3–30 mmollitre of sugars. The variation coefficients were 3.2% for the GOD electrode and 1.6% for the AG electrode in measuring the mixture of glucose and maltose. The results made by the dual-electrode system agreed well with those made by Fehling titration. During 20 daysʹ determination, the activities of the enzyme electrodes showed no decay.
Journal title :
Food Chemistry
Serial Year :
1995
Journal title :
Food Chemistry
Record number :
1946283
Link To Document :
بازگشت