Title of article :
Automatic determination of nickel in foods by flame atomic absorption spectrometry
Author/Authors :
Yebra، نويسنده , , M.C. and Cancela، نويسنده , , S. and Cespَn، نويسنده , , R.M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
5
From page :
774
To page :
778
Abstract :
A new sensitive and low cost flow injection method that combines acid extraction, preconcentration and flame atomic absorption spectrometric determination of nickel in food samples at μg/g levels is described. The dynamic acid extraction step was carried out by using a continuous ultrasound-assisted extraction system. The acid extract was preconcentrated on-line on a minicolumn packed with a chelating resin (Serdolit Che, with iminodiacetic groups) and nickel was eluted with diluted hydrochloric acid, being continuously monitored by flame atomic absorption spectrometry. An experimental design (Plackett-Burman 26 × 3/16) is used to optimise the methodology proposed. The method allowed a total sampling frequency of 13–28 samples per hour. Good precision of the whole procedure (1.9–3.6% expressed as relative standard deviation) and a detection limit of 0.12 μg/g, for 60 mg of sample were achieved. The method was successfully applied to the determination of trace amounts of nickel in food samples.
Keywords :
Ultrasound-assisted acid extraction , Minicolumn preconcentration , Flow injection analysis , nickel , Food analysis
Journal title :
Food Chemistry
Serial Year :
2008
Journal title :
Food Chemistry
Record number :
1956849
Link To Document :
بازگشت