• Title of article

    A non-plasticized chitosan based solid state electrode for flow injection analysis of glutamate in food samples

  • Author/Authors

    Isa، نويسنده , , Illyas Md and Ab Ghani، نويسنده , , Sulaiman، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    4
  • From page
    756
  • To page
    759
  • Abstract
    This report describes the development of solid state glutamate sensor based on chitosan. The electrode has linearity range of 0.01–1 mM glutamate with detection limit of 0.008 mM. The presence of other conventional food additives at physiological level does not interfere. The interfering effect is, however, minimized through prior dilution of the sample. Recovery values of 88.9–99.2% are obtained throughout. The proposed electrode has been applied in flow injection analysis (FIA) glutamate in food samples. In the validation experiment, the proposed electrode is found to be comparable with the standard method.
  • Keywords
    glutamate , Non–plasticized , Solid state electrode , Flow injection analysis , Graphite pencil rod
  • Journal title
    Food Chemistry
  • Serial Year
    2009
  • Journal title
    Food Chemistry
  • Record number

    1957450