• Title of article

    Flow injection analysis of BHA by NiHCF modified electrode

  • Author/Authors

    S.J. Richard Prabakar، نويسنده , , S.J. and Sriman Narayanan، نويسنده , , S.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    7
  • From page
    449
  • To page
    455
  • Abstract
    Flow injection method is described for the determination of the antioxidant tert-butylhydroxy anisole (BHA) based on its catalytic oxidation at a nickel hexacyanoferrate (NiHCF) surface modified graphite wax composite electrode fabricated using a new approach. The electrochemical characteristics of the modified electrode were studied using cyclic voltammetry. The voltammetric response of BHA at the modified electrode showed current densities remarkably higher than the bare graphite electrode and occurred at a reduced over voltage of 200 mV. Linear calibration graphs were obtained in the range of 1.2 × 10−6–1.07 × 10−3 M. A detection limit of 6 × 10−7 M was obtained with a correlation coefficient of 0.9983 based on S/N = 3. Reliable results were obtained by applying the proposed flow injection method to determine BHA spiked in dehydrated potato flakes. Results suggest that the developed method can be effectively employed for the determination of BHA from food samples.
  • Keywords
    Flow injection analysis , Amperometric determination , Electrocatalytic oxidation , Nickel hexacyanoferrate , Butylated hydroxy anisole
  • Journal title
    Food Chemistry
  • Serial Year
    2010
  • Journal title
    Food Chemistry
  • Record number