Title of article :
Certification of the reference material of water content in water saturated 1-octanol by Karl Fischer coulometry, Karl Fischer volumetry and quantitative nuclear magnetic resonance
Author/Authors :
Wang، نويسنده , , Haifeng and Ma، نويسنده , , Kang and Zhang، نويسنده , , Wei and Li، نويسنده , , Jia and Sun، نويسنده , , Guohua H. Li، نويسنده , , Hongmei، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
5
From page :
2362
To page :
2366
Abstract :
Certified reference materials (CRMs) of water content are widely used in the calibration and validation of Karl Fischer coulometry and volumetry. In this study, the water content of the water saturated 1-octanol (WSO) CRM was certified by Karl Fischer coulometry, volumetry and quantitative nuclear magnetic resonance (Q NMR). The water content recovery by coulometry was 99.76% with a diaphragm-less electrode and Coulomat AG anolyte. The relative bias between the coulometry and volumetry results was 0.06%. In Q NMR, the water content of WSO is traceable to the International System (SI) of units through the purity of internal standard. The relative bias of water content in WSO between Q NMR and volumetry was 0.50%. The consistency of results for these three independent methods improves the accuracy of the certification of the RM. The certified water content of the WSO CRM was 4.76% with an expanded uncertainty of 0.09%.
Keywords :
Karl Fischer titration , certified reference material , Quantitative nuclear magnetic resonance , water content
Journal title :
Food Chemistry
Serial Year :
2012