Title of article
Development and validation of the high performance liquid chromatography–ion exclusion method for detection of lactic acid in milk
Author/Authors
Milagres، نويسنده , , Maria Patricia and Brandمo، نويسنده , , Sebastiمo César Cardoso and Magalhمes، نويسنده , , Mirella Araujo and Minim، نويسنده , , Valéria Paula Rodrigues and Minim، نويسنده , , Luis Antonio، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
5
From page
1078
To page
1082
Abstract
The present work reports the development and validation of an analytical method for lactic acid quantification in milk by ion exclusion–HPLC with UV detection. This method showed good precision (variation coefficient <7.0%) and good accuracy (97.6–99.6%). Calibration curves were linear (R2 > 0.996) at the tested range of 5.0–25.0 mmol L−1; and the detection and quantification limits were 0.5 and 1.0 mmol L−1, respectively. Analysis of lactic acid in fermented milk (FM) or NaOH neutralised fermented milk (FNM) was performed using samples obtained during 9 h of fermentation. An ANOVA analysis combined with the Tukey test was used to show that the method was insensitive to the differences between FM and FNM. The proposed method therefore proved to be a useful quality control tool in laboratories because it was able to determine milk acidity even in adulterated samples.
Keywords
Ion exclusion , Method development , milk , lactic acid , High-performance liquid chromatography , Method validation
Journal title
Food Chemistry
Serial Year
2012
Journal title
Food Chemistry
Record number
1970513
Link To Document