Title of article
Improved glucose electrochemical biosensor by appropriate immobilization of nano-ZnO
Author/Authors
Lei، نويسنده , , Yang and Yan، نويسنده , , Xiaoqin and Zhao، نويسنده , , Jing and Liu، نويسنده , , Xi and Song، نويسنده , , Yu and Luo، نويسنده , , Ning and Zhang، نويسنده , , Yue، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
5
From page
168
To page
172
Abstract
We constructed the transferred ZnO biosensor and the grown ZnO biosensor by two different nano-ZnO immobilization approaches. And the influence of different assembly processes on the biosensor performance has been systematically investigated and compared. An enhanced sensitivity of the grown ZnO biosensor is found to be 52% higher than that of the transferred ZnO biosensor. Correspondingly, the other properties are also better in the grown ZnO biosensor, including the response time, the detection limit and the linear range. These results are well consistent with the fact that more glucose oxidase is immobilized on the well-aligned ZnO arrays, which have higher specific surface area and more direct electron communication path, in the grown sensor than the randomly distributed and stacked ZnO nanorods in the transferred sensor. The nano-ZnO grown directly has been demonstrated more desirable for enzymatic immobilization and signal transduction in the high performance biosensors.
Keywords
Glucose oxidase , ZnO nanorod , Electrochemistry , Biosensor
Journal title
Colloids and Surfaces B Biointerfaces
Serial Year
2011
Journal title
Colloids and Surfaces B Biointerfaces
Record number
1972228
Link To Document